FED (E2) 12/2014
page 16/55
3.4 Location of installation and ambient conditions
Set up the drying and heating oven FED on an even and non-flammable surface, free from vibration and
in a well-ventilated, dry location and align it using a spirit level. The site of installation must be capable of
supporting the unit’s weight (see technical data, chap. 12.4). The chambers are designed for setting up
inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the unit.
∅
Do NOT set up units in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation: +18 °C up to +40 °C. At elevated ambient
temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient
temperature of +25 °C to which the specified technical data relate. For other ambient
conditions, deviations from the indicated data are possible.
•
Permissible ambient humidity: 70 % r.H. max., non-condensing.
•
Installation height: max. 2000 m /
6562 ft.
above sea level.
When placing several units of the same size side by side, maintain a minimum distance of 250 mm
between each unit. Wall distances: rear 100 mm, sides 160 mm. Spacing above the unit of at least 100
mm must also be accounted for.
Two devices up to size 115l can be piled on top of each other. For this purpose, place rubber pads under
all four feet of the upper unit to prevent the device from slipping.
CAUTION
Sliding or tilting of the upper unit.
Damage to the units.
When stacking, place rubber pads under all four feet of the upper unit.
To completely separate the unit from the power supply, you must disconnect the power plug. Install the
unit in a way that the power plug is easily accessible and can be easily pulled in case of danger.
Do not install or operate the drying and heating oven FED in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT operate the unit in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit..