FED (E2) 12/2014
page 13/55
Due to the special demands of the Medical Device Directive (MDD), these ovens are not
qualified for sterilization of medical devices as defined by the directive 93/42/EWG.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 9) are part of the intended use.
The charging material shall not contain any corrosive ingredients that may damage the
machine components. Such ingredients include in particular acids and halides. Any corrosive
damage caused by such ingredients is excluded from liability by BINDER GmbH.
2.
Unit description
BINDER drying and heating ovens FED are equipped with an electronic PID-controller with digital display.
The temperature is indicated with an accuracy of one degree.
BINDER drying and heating ovens with forced convection FED are heated electrically and are ventilated
by fan-assisted, forced-air circulation. They FED are equipped with a temperature safety device
according to DIN12880 (chap. 7).
The APT.line™ preheating chamber system guarantees high level of spatial and time-based temperature
precision, thanks to the direct and distributed air circulation into the interior. The fan supports exact
attainment and maintenance of the desired temperature accuracy.
The inner chamber, the pre-heating chamber and the inside of the doors are all made of stainless steel
V2A (German material no. 1.4301, US equivalent AISI 304). When operating the unit at temperatures
above 150 °C, the impact
of the oxygen in the air may cause discoloration of the metallic surfaces
(yellowish-brown or blue) by natural oxidation processes. These colorations are harmless and will in no
way impair the function or quality of the unit. The housing is RAL 7035 powder-coated. All corners and
edges are also completely coated.
All unit functions are easy and comfortable to use thanks to their clear arrangement. Major features are
easy cleaning of all unit parts and avoidance of undesired contamination.
BINDER drying and heating ovens FED are equipped with a serial interface RS 422 for computer
communication, e.g. via the communication software APT-COM™ 3 DataControlSystem (option, chap.
8.1). For further options, see chap. 12.5.
The model FED 720 is equipped with
four castors. Both
front
castors
can be locked by brakes.
The unit can be operated in a temperature range of 5 °C /
9 °F
above room temperature up to 300 °C /
572 °F
.