CB (E7) 06/2018
Page 158/174
25. Technical description
25.1 Factory calibration and adjustment
This chamber was calibrated and adjusted in the factory. Calibration and adjustment were performed
using standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER
(certified since December 1996 by TÜV CERT). All test equipment used is subject to the administration of
measurement and test equipment that is also a constituent part of the BINDER QM DIN EN ISO 9001
systems. They are controlled and calibrated to a DKD-Standard at regular intervals.
A record of this calibration and adjustment is part of the BINDER test certificate of the chamber.
Adjustment in factory:
•
Temperature
: 37 °C /
98.6°F
measured in the center of the usable volume
•
CO
2
: 0 vol.-% CO
2
(100 vol.-% N
2
) and 5 vol.-% CO
2
(analyzed test gas directly exposed to the sensor
head)
•
O
2
(chamber with O
2
control): 0 vol.-% O
2
(100 vol.-% N
2
, analyzed test gas directly exposed to the
sensor head) and 20.9 vol.-% O
2
(ambient air).
•
Humidity
(chamber with active humidification): 93 % r.h.
Repeated calibrations are recommended in periods of 12 months.
BINDER service uses an electronic measuring and display device for temperature traceable to an
acknowledged standards/calibration institution (DKD or PTB for Germany) with valid calibration certificate.
Test gases with an analyzed concentration serve to calibrate the sensor systems for CO
2
and O
2
(chamber with O
2
control). The sensor heads are exposed directly to the test gas.
25.2 Over current protection
The CB is protected by a chamber-protection against over current, accessible from the outside. It is
located at the rear of the chamber below the strain relief of the power cord. The fuse holder is equipped
with a fuse clip 5mm x 20 mm (cUL-Version 6.3 x32 mm). Replace this fuse only with a substitute of the
same ratings. Refer to the technical data of the respective device type. If this fuse is blown, please inform
an electronic engineer or BINDER service.
Summary of Contents for CB 170
Page 165: ...CB E7 06 2018 Page 165 174 25 9 Dimensions CB 170 Dimensions in mm...
Page 167: ...CB E7 06 2018 Page 167 174...
Page 168: ...CB E7 06 2018 Page 168 174 27 Product registration...
Page 175: ......