CB (E7) 06/2018
Page 86/174
10.6.4 Setpoint entry
•
Select the field “Temperature” and enter the desired temperature setpoint.
Setting range: 20 °C up to 60 °C.
Confirm entry with
Confirm
icon. The controller returns to the section view.
•
Select the field “CO2” and enter the desired CO
2
setpoint.
Setting range: 0 vol.-% up to 20 vol.-%.
Confirm entry with
Confirm
icon. The controller returns to the section view.
Chamber with O
2
control:
•
Select the field “O2” and enter the desired O
2
setpoint.
Setting range with regular hypoxic control range: 0.2 vol.-% up to 20 vol.-%.
Setting range with alternative control range: 5 vol.-% up to 95 vol.-%.
Confirm entry with
Confirm
icon. The controller returns to the section view.
Chamber with active humidification:
•
Select the field “Humidity” and enter the desired humidity setpoint.
Setting range: 60% r.h. up to 93% r.h.
Confirm entry with
Confirm
icon. The controller returns to the section view.
10.6.5 Special controller functions via operation lines
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Chamber with O
2
control:
Depending on the required range of O
2
concentration, the type of gas which is
required may vary (chap. 6.6). Therefore, the pressure alarms for O
2
and N
2
can be deactivated
individually with active O
2
control.
For chambers equipped with the optional interior socket (chap. 18.3) you can turn on and off the voltage
of the interior socket via the controller.
•
Operation line “Idle mode” activates / deactivates the operating mode “Idle mode”.
•
Operation line “O2 pressure alarm off” activates / deactivates the O
2
pressure alarm (chamber with O
2
control)
•
Operation line “N2 pressure alarm off” activates / deactivates the N
2
pressure alarm (chamber with O
2
control)
•
Operation line “Interior socket” activates / deactivates interior socket (option, available via BINDER
INDIVIDUAL Customized Solutions)
The other operation lines are without function.
Select the desired program and section. You can set the operation lines in the “Functions on/off” field.
For details please refer to chap. 9.7.3.
Summary of Contents for CB 170
Page 165: ...CB E7 06 2018 Page 165 174 25 9 Dimensions CB 170 Dimensions in mm...
Page 167: ...CB E7 06 2018 Page 167 174...
Page 168: ...CB E7 06 2018 Page 168 174 27 Product registration...
Page 175: ......