control is deactivated and will only be activated after confirming the information
message “Successful sterilization” and when the temperature has cooled down to a value
below < 80 °C.
22. Maintenance and service
22.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of death.
∅
The chamber must NOT become wet during operation or maintenance works.
∅
Do NOT remove the rear panel of the chamber.
Before conducting maintenance work, turn off the chamber at the main power switch
and disconnect the power plug.
Any maintenance work must be conducted by licensed electricians or experts
authorized by BINDER.
Ensure regular maintenance work is performed at least once a year and that the legal requirements are
met regarding the qualifications of service personnel, scope of testing and documentation.
The warranty becomes void if maintenance work is conducted by non-authorized personnel.
Chamber with active humidification:
Have conducted regular maintenance work on the steam humidifier
at least once a year. The operating behavior and the maintenance intervals of the humidifier essentially
depend on the available water quality and the amount of steam produced in the meantime.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.
We recommend taking out a maintenance agreement. Please consult BINDER Service.
BINDER telephone hotline:
+49 (0) 7462 2005 555
BINDER fax hotline:
+49 (0) 7462 2005 93555
BINDER e-mail hotline:
BINDER service hotline USA:
+1 866 885 9794 or +1 631 224 4340 x3 (toll-free in the USA)
BINDER service hotline Asia Pacific:
+852 390 705 04 or +852 390 705 03
BINDER service hotline Russia and CIS
+7 495 988 15 16
BINDER Internet website
http://www.binder-world.com
BINDER address
BINDER GmbH, post office box 102,
78502 Tuttlingen, Germany
International customers, please contact your local BINDER distributor.
The gas sensors are especially adjusted for the specific incubator. When exchanging a
sensor, you must repeat control adjustment for CO
2
and O
2
(chamber with O
2
control).