36
NOTE I
At 80 MHz end 800 MHz. the higher frequency range applies
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the
location in which the model is used exceeds the applicable RF compliance level above, the
model should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the model .
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the model
The model is intended for use in electromagnetic environment in which radiated RF
disturbances is controlled. The customer or the user of the model can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment(transmitters )and the model is recommended below,
according to the maximum output power of the communications equipment.