35
Guidance & Declaration - Electromagnetic immunity
The model is intended for use in the electromagnetic environment specified below. The
customer or the user of the model should assure that it is used in such an environment.
Immunity test IEC 60601
test level
ComplianceI
level
Electromagnetic environment - guidance
Conducted
RF IEC
61000-4-
6
Radiated RF
IEC 61000-4-
3
3 Vrms
150kHz to
80 MHz 3
V/m
80 MHz to
2.5 GHz
3 V
3 V/m
Portable and mobile RF
communications equipment should be used
no closer to any part of the model, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
Transmitter
Recommended separation distance
d=[3,5/V1]XP1/2
d=1.2xP1/280 MHz to 800 MHz
d=2.3xP1/2 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter
In watts (W) according to the transmitter
manufacturer and d Is the recommended
separation distance in meters
(m) Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey," should be less than the compliance
level in each frequency range b.
Interference may occur In the vicinity of
equipment marked with the following symbol: