and verify the suitability and effectiveness. *Fractional pre-vacuum procedure = steam
sterilization with repetitive prevacuum. The procedure used here is to perform steam
sterilization through three prevacuums.
7.9 Storage
•
Store in a clean, dry, ventilated, non-corrosive atmosphere with a relative humidity of 10% to
93%, an atmospheric pressure of 70 kPa to 106KPa, and a temperature of -20°C to +55°C;
•
After sterilization, the product should be packaged in a medical sterilization bag and stored
in a special storage cabinet. The storage time should not exceed 7 days. If it is exceeded, it
should be reprocessed before use.
Notes:
1. The storage environment should be clean and must be disinfected regularly;
2. Product storage must be batched and marked and recorded.
7.10 Transportation
•
Prevent excessive shock and vibration during transportation, and handle with care;
•
It should not be mixed with dangerous goods during transportation.
•
Avoid exposure to sun or rain or snow during transportation.
•
The cleaning and disinfection of main unit and the measuring wire are as follows.
•
Before each use, wipe the surface of the machine and the measuring wire with a soft cloth or
paper towel soaked in 75% medical alcohol. Repeat the wipe for at least 3 times.
•
After each use, wipe the surface of the device and the measuring wire with a soft cloth
soaked in clean water (distilled or deionized water) or a clean disposable wipe. Repeat the
wipe for at least 3 times.