24
Sterilize it after disinfection and use: temperature ≥ 90 ° C, time ≥ 1 min or A0 ≥ 600 (b) For
the disinfection here, the temperature is 93 ° C, the time is 2.5 min, and A0>3000
5. Only distilled or deionized water with a small amount of microorganisms
(<10 cfu/ml) can be used for all rinsing steps. (For example, pure water that is in accordance
with the European Pharmacopoeia or the United States Pharmacopoeia).
6. After cleaning, the chemical residue should be less than 10mg / L.
7. The air used for drying must be filtered by HEPA.
8. Regularly repair and inspect the disinfector.
7.5 Drying
If your cleaning and disinfection process does not have an automatic drying function, dry it
after cleaning and disinfection.
Methods:
1. Spread a clean white paper (white cloth) on the flat table, point the product against the
white paper (white cloth), and then dry the product with filtered dry compressed air (maximum
pressure 3 bar). Until no liquid is sprayed onto the white paper (white cloth), the product
drying is completed.
2. It can also be dried directly in a medical drying cabinet (or oven). The recommended drying
temperature is 80°C~120°C and the time should be 15~40 minutes.
Note:
1. The drying of product must be performed in a clean place.
2. The drying temperature should not exceed 138 °C;
3. The equipment used should be inspected and maintained regularly.
7.6 Inspection and maintenance
1. Check the product. If there is still visible stain on the product after cleaning/ disinfection, the
entire cleaning/disinfection process must be repeated.
2. Check the product. If it is obviously damaged, smashed, detached, corroded or ruptured, it
must be disposed and not allowed to continue to be used.
3. Check the product. If the accessories are found to be damaged, please replace it before
use. The new accessories for replacement must be cleaned, disinfected and dried.
4.If the service time (number of times) of the product reaches the specified service
life (number of times), please replace it in time.