background image

                       

BestScope International Limited

 

-

 

dust affect bulb brightness and service life, and insert fully the contact pins into the bulb socket. 
6.3.6 Close the lamp housing cover, and screw the knurled thumb screw. 

After working for above 10 hours continuously, 

better cut off the microscope about 30 minutes. 
6.4 Replacement of fuse 
6.4.1 Cut off power of microscope, and pull out the 
plug. 
6.4.2 Unscrew fuse cap in the back of base, take out 
old fuse. 
6.4.3 Replace a new fuse, then screw the fuse cap. 

7. Trouble shooting 

In the period of using this series microscope, if there is any trouble occurs, please referring to the 
following sheet listed some common troubleshooting resolve them.

 

Trouble 

Causation 

Remedy 

Switch on but bulb dark

 

Plug is unreliable 

Plug in again       

Bulb is broken 

Change bulb 

Fuse is broken 

Change fuse 

Bulb is flickering or 
brightness is unsteady 

Bulb is unstable 

Insert it again 

Bulb is broken 

Replacing bulb 

Brightness  of  view  field 

isn’t  e

nough  or  is 

Uneven 

Bulb  specification  doesn’t  meet 

the requirement 

Replacing bulb 

Brightness isn’t adjusted

 

correctly 

Adjust rotation 
potentiometer 

Objective isn’t in correct

 

position 

Make  the  objective  in  correct 
position 

The  size  of  iris  aperture  is  too 
small 

Adjust the size of iris aperture 

Brightness  of  view  field 

isn’t  enough  or  is 

Uneven 

Lens(objective,eyepiece, 
condenser,  light  collector)  has 
dust 

Clean it

 

Position of condenser is too low 

Higher condenser 

Image isn’t clear

 

  (contrast  or  definition 

isn’t enough)

 

Cover  glass  of  specimen  doesn’t 

meet the requirement 

Use  required  thickness  cover  glass 
(0.17mm) 

Cover  glass  of  specimen  isn’t  in 

up direction 

Place specimen correctly 

Surface of objective lens 
isdirty  (especially  it  is  easy  for 
the front lens of 40X objective to 
dip in immersion oil) 

Clean it 

Immersion oil isn’t used for 100X 

objective (oil) 

Use immersion oil

 

Immersion  oil  doesn’t  meet  the 

requirement 

Use immersion oil 
supplied by us 

There is bubble in immersion oil 

Clear the bubble way

 

Summary of Contents for BS-2036 Series

Page 1: ...ical microscope To ensure the safety obtain optimum performance and to familiarize you fully with the use of this microscope it is recommended strongly that you study this manual thoroughly before usi...

Page 2: ...enance 6 7 Troubleshooting 7 8 Outfits 8 BS 2036 series biological microscopes are designed for college teaching Medical and clinical identification The microscope with modern design steady structure...

Page 3: ...cular Head Trinocular Head Power Input Fuse F L M N A Diopter Adjustment Ring B Nosepiece C Clamp Handle D Handle of Iris Aperture Diaphragm E Field Diaphragm Ring F Body Handle G Head Fixing Screw H...

Page 4: ...e Field Resolving Power Working Distance 4X 0 10 5mm 2 8 m 16mm 10X 0 25 2mm 1 1 m 2mm 20X S Optional 0 40 1mm 0 69 m 1 1mm 40X S 0 65 0 5mm 0 42 m 0 66mm 100X Oil S 1 25 0 2mm 0 22 m 0 42mm BS 2036C...

Page 5: ...spherical System 2 3 9 Filter Built in blue filer 2 3 10 Electric components Input voltage AC85 265V LED AC170 270V 50 60Hz Output voltage DC1 2 6V LED DC0 4 3 5V 6V 20W halogen lamp Rotation potenti...

Page 6: ...phragm Adjust condenser condenser focusing screw M and aperture Aperture diaphragm lever D Observing 1 Turn power switch adjust light Turn on Switch L and turn Potentiometer K 2 Put the sample Put the...

Page 7: ...preparation is suggested to clean it 6 1 4 Plastic should be cleaned by soft cloth with clear water 6 2 Environment of using and placing 6 2 1 Microscope should be used and placed in a cool dry non du...

Page 8: ...t again Bulb is broken Replacing bulb Brightness of view field isn t enough or is Uneven Bulb specification doesn t meet the requirement Replacing bulb Brightness isn t adjusted correctly Adjust rotat...

Page 9: ...en correctly Cover glass doesn t meet the requirement Use required thickness cover glass 0 17mm Image observed by two eyes aren t in superposition entirely Interpupilary distance isn t adjusted correc...

Page 10: ...tribution 20 80 Trinocular Head For Infinite System Inclined 30 Rotatable 360 Interpupilary Distance 48 75mm Light Distribution 20 80 Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76...

Reviews: