5. Image Capture 

5.1 Installing 

Connect the C-mount with CCD camera or connect camera with camera adapter, then connect it 

with c-mount, finally put it into microscope. 

5.2 Using 

First get a clear image from eyepiece, then put out lever on the side of trinocular head, and collect 

image with camera. Clear image should be in screen. Adjust B14 fine focusing knobs to get it clear if 

image is not clear. 

6. Maintenance 

6.1 Clean microscope 

6.1.1 Don’t touch the lens with hand, Dust on lens should be cleaned by soft brush or absorbent 

cotton or cleaned by absorbent cotton, lens paper with the mixture of alcohol and ether 
(proportion 1:4). 
6.1.2 Alcohol and ether all are burnt early, please take them away from fire. Be careful for turn on 
and off power. 

6.1.3 Don’t clean painted metal and galvanizing metal with organic solvent such as alcohol, ether or 

the mixture of the both. Silicon cloth or soft cleaning preparation is suggested to clean it. 
6.1.4 Plastic should be cleaned by soft cloth with clear water. 
 
6.2 Environment of using and placing 
6.2.1 Microscope should be used and placed in a cool, dry, non-dust, non-shake and non-corrosive 
gases environment. 
6.2.2 Microscope should be used in environment of indoor temperature 0°-40°C and maximum 
relative humidity 85%. 
6.2.3 Removing equipment is suggested to be installed when microscope used in heavy humidity 
area to avoid fungus and mist damage instrument. 
6.2.4 Please pay attention to prevent microscope from violent shake and vibration in application 

and in carrying. Don’t drag it on the surface of worktable to avoid damage to microscope and 

worktable. 
 
6.3 Replacement of bulb 
6.3.1 Turn off power, and pull out plug. 
6.3.2 Wait the bulb become cool. 

 Please be sure that the bulb is cool, then follow 

by the nest operations. 
6.3.3 Lay aside the microscope reliably, unscrew the 
knurled thumb screw of the lamp housing cover on 
the underside of base. 
6.3.4 Pull over the lamp housing cover. 
6.3.5 Pull out the bulb should be replaced, hold a new bulb with silk cloth to avoid fingerprint and 

Summary of Contents for BS-2036 Series

Page 1: ...ical microscope To ensure the safety obtain optimum performance and to familiarize you fully with the use of this microscope it is recommended strongly that you study this manual thoroughly before usi...

Page 2: ...enance 6 7 Troubleshooting 7 8 Outfits 8 BS 2036 series biological microscopes are designed for college teaching Medical and clinical identification The microscope with modern design steady structure...

Page 3: ...cular Head Trinocular Head Power Input Fuse F L M N A Diopter Adjustment Ring B Nosepiece C Clamp Handle D Handle of Iris Aperture Diaphragm E Field Diaphragm Ring F Body Handle G Head Fixing Screw H...

Page 4: ...e Field Resolving Power Working Distance 4X 0 10 5mm 2 8 m 16mm 10X 0 25 2mm 1 1 m 2mm 20X S Optional 0 40 1mm 0 69 m 1 1mm 40X S 0 65 0 5mm 0 42 m 0 66mm 100X Oil S 1 25 0 2mm 0 22 m 0 42mm BS 2036C...

Page 5: ...spherical System 2 3 9 Filter Built in blue filer 2 3 10 Electric components Input voltage AC85 265V LED AC170 270V 50 60Hz Output voltage DC1 2 6V LED DC0 4 3 5V 6V 20W halogen lamp Rotation potenti...

Page 6: ...phragm Adjust condenser condenser focusing screw M and aperture Aperture diaphragm lever D Observing 1 Turn power switch adjust light Turn on Switch L and turn Potentiometer K 2 Put the sample Put the...

Page 7: ...preparation is suggested to clean it 6 1 4 Plastic should be cleaned by soft cloth with clear water 6 2 Environment of using and placing 6 2 1 Microscope should be used and placed in a cool dry non du...

Page 8: ...t again Bulb is broken Replacing bulb Brightness of view field isn t enough or is Uneven Bulb specification doesn t meet the requirement Replacing bulb Brightness isn t adjusted correctly Adjust rotat...

Page 9: ...en correctly Cover glass doesn t meet the requirement Use required thickness cover glass 0 17mm Image observed by two eyes aren t in superposition entirely Interpupilary distance isn t adjusted correc...

Page 10: ...tribution 20 80 Trinocular Head For Infinite System Inclined 30 Rotatable 360 Interpupilary Distance 48 75mm Light Distribution 20 80 Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76...

Reviews: