BestScope International Limited

 

-

Condenser 

screw 

Condenser 

installing ring 

Condenser 

2.3 The other specification 

2.3.1 Mechanical tube length: 160mm 

2.3.2 Conjugate distance: Infinity (Non- Infinity: 195mm) 

2.3.3 Head: Seidentopf Binocular or Trinocular, Inclined 30°, Rotatable 360°,    Anti-fungal                                                                                   

systems. Interpupillary Adjustable Distance Is 48-75mm, Diopter adjustable range ±5. 

2.3.4 Nosepiece: Quadplex nosepiece 

2.3.5 Mechanical stage: Size 145mm×140mm, X-Y travel 76mm×52mm 

2.3.6 Focusing systems: Coaxial Coarse and Fine Focusing Knobs, Coarse stroke 26mm, 

Fine division 2µm, Condenser up-down range 22mm 

2.3.7 Condenser: Abbe condenser, N.A. 1.25, Adjustable aperture, Aperture center can be 

adjustable. 

2.3.8 Illumination: Non-spherical System   

2.3.9 Filter: Built in blue filer 

2.3.10 Electric components: Input voltage AC85-265V/LED AC170-270V, 50/60Hz 

                                        Output voltage DC1.2-6V/LED, DC0.4-3.5V 6V/20W halogen lamp 

                                        Rotation potentiometer with power switch,   

                                       

Fuse    2A ф5×20

 

3.

 

Installation 

Please  clean  the  operation  desk  before  installation.  Put 
out the microscope of the carton and put it on the desk.   

Make  sure  the  supply  voltage  meets  the  instrument’s 

requirement and the power switch is off. 
 
Installation Instruction Fig.: 
1.

 

Turn the binocular (trinocular) head to working 
position; 

2.

 

Put out the dust cover of the eyepiece tube; 

3.

 

Insert into the eyepiece; 

4.

 

Install Abbe Condenser 

 
3.1 Eyepiece Tube: 
Loose  the  Head  Fixing  Screw  G,  turn  the  tube  to 
observing position, then tighten the Screw G. 
 
3.2 Put out the eyepiece dust cover. 
 
3.3 Eyepiece 

Put out the eyepiece of the carton, and insert it into the tube. Please don’t touch the lens of the 

1

 

2

 

3

 

4

 

Summary of Contents for BS-2036 Series

Page 1: ...ical microscope To ensure the safety obtain optimum performance and to familiarize you fully with the use of this microscope it is recommended strongly that you study this manual thoroughly before usi...

Page 2: ...enance 6 7 Troubleshooting 7 8 Outfits 8 BS 2036 series biological microscopes are designed for college teaching Medical and clinical identification The microscope with modern design steady structure...

Page 3: ...cular Head Trinocular Head Power Input Fuse F L M N A Diopter Adjustment Ring B Nosepiece C Clamp Handle D Handle of Iris Aperture Diaphragm E Field Diaphragm Ring F Body Handle G Head Fixing Screw H...

Page 4: ...e Field Resolving Power Working Distance 4X 0 10 5mm 2 8 m 16mm 10X 0 25 2mm 1 1 m 2mm 20X S Optional 0 40 1mm 0 69 m 1 1mm 40X S 0 65 0 5mm 0 42 m 0 66mm 100X Oil S 1 25 0 2mm 0 22 m 0 42mm BS 2036C...

Page 5: ...spherical System 2 3 9 Filter Built in blue filer 2 3 10 Electric components Input voltage AC85 265V LED AC170 270V 50 60Hz Output voltage DC1 2 6V LED DC0 4 3 5V 6V 20W halogen lamp Rotation potenti...

Page 6: ...phragm Adjust condenser condenser focusing screw M and aperture Aperture diaphragm lever D Observing 1 Turn power switch adjust light Turn on Switch L and turn Potentiometer K 2 Put the sample Put the...

Page 7: ...preparation is suggested to clean it 6 1 4 Plastic should be cleaned by soft cloth with clear water 6 2 Environment of using and placing 6 2 1 Microscope should be used and placed in a cool dry non du...

Page 8: ...t again Bulb is broken Replacing bulb Brightness of view field isn t enough or is Uneven Bulb specification doesn t meet the requirement Replacing bulb Brightness isn t adjusted correctly Adjust rotat...

Page 9: ...en correctly Cover glass doesn t meet the requirement Use required thickness cover glass 0 17mm Image observed by two eyes aren t in superposition entirely Interpupilary distance isn t adjusted correc...

Page 10: ...tribution 20 80 Trinocular Head For Infinite System Inclined 30 Rotatable 360 Interpupilary Distance 48 75mm Light Distribution 20 80 Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76...

Reviews: