Trouble 

Causation 

Remedy 

Size of iris aperture isn’t proper

 

Adjust the size of iris aperture 

Position of condenser is too low 

Readjust the position of 
condenser 

One side of image is 
dark or image is 
moving as focusing   

Objective isn’t in correct position

 

Make the objective in 
correct position 

Specimen isn’t placed correctly

 

Place specimen levelly on stage and 
clip it with clamp 

Objective touches 
specimen  as  changing 
low  times  objective  to 
high times objective 

Cover  glass  of  specimen  isn’t  in 

up direction 

Place specimen correctly

 

Cover  glass  doesn’t  meet  the 

requirement 

Use required thickness 
cover glass (0.17mm) 

Image  observed  by  two 

eyes aren’t in

 

superposition entirely. 

Interpupilary distance isn’t 

adjusted correctly   

Adjust interpupilary   
distance according to two eyes 

It  is  easy  for  eyes  to  be 
tired during observing   

Diopter isn’t adjusted correctly

 

Readjust diopter 

8. Outfits

 

Items 

Specification 

BS-2036

BS-2036

BS-2036

BS-2036

Eyepiece 

WF 10×-18 mm 

 

 

 

 

WF 10×-20mm 

 

 

 

 

WF16×-13mm 

 

 

 

 

Reticule Eyepiece WF 10×-18 mm   
(Reticule 0.1mm) 

 

 

 

 

Reticule Eyepiece WF 10×-20mm   
(Reticule 0.1mm) 

 

 

 

 

Achromatic 
Objective 

4× 

 

 

 

 

10× 

 

 

 

 

20× 

 

 

 

 

40×(S) 

 

 

 

 

100×/1.25 (Oil) (S)   

 

 

 

 

Plan 
Objective 

4× 

 

 

 

 

10× 

 

 

 

 

40×/0.65 (S) 

 

 

 

 

100×/1.25 (Oil) (S) 

 

 

 

 

Infinite 
E-Plan   
Objectives 

4× 

 

 

 

 

10× 

 

 

 

 

20× 

 

 

 

 

40× (S) 

 

 

 

 

100× (Oil) (S) 

 

 

 

 

Infinite Plan 
Achromatic 
Objective 

4× 

 

 

 

 

10× 

 

 

 

 

20× 

 

 

 

 

40× (S) 

 

 

 

 

60x(S) 

 

 

 

 

Summary of Contents for BS-2036 Series

Page 1: ...ical microscope To ensure the safety obtain optimum performance and to familiarize you fully with the use of this microscope it is recommended strongly that you study this manual thoroughly before usi...

Page 2: ...enance 6 7 Troubleshooting 7 8 Outfits 8 BS 2036 series biological microscopes are designed for college teaching Medical and clinical identification The microscope with modern design steady structure...

Page 3: ...cular Head Trinocular Head Power Input Fuse F L M N A Diopter Adjustment Ring B Nosepiece C Clamp Handle D Handle of Iris Aperture Diaphragm E Field Diaphragm Ring F Body Handle G Head Fixing Screw H...

Page 4: ...e Field Resolving Power Working Distance 4X 0 10 5mm 2 8 m 16mm 10X 0 25 2mm 1 1 m 2mm 20X S Optional 0 40 1mm 0 69 m 1 1mm 40X S 0 65 0 5mm 0 42 m 0 66mm 100X Oil S 1 25 0 2mm 0 22 m 0 42mm BS 2036C...

Page 5: ...spherical System 2 3 9 Filter Built in blue filer 2 3 10 Electric components Input voltage AC85 265V LED AC170 270V 50 60Hz Output voltage DC1 2 6V LED DC0 4 3 5V 6V 20W halogen lamp Rotation potenti...

Page 6: ...phragm Adjust condenser condenser focusing screw M and aperture Aperture diaphragm lever D Observing 1 Turn power switch adjust light Turn on Switch L and turn Potentiometer K 2 Put the sample Put the...

Page 7: ...preparation is suggested to clean it 6 1 4 Plastic should be cleaned by soft cloth with clear water 6 2 Environment of using and placing 6 2 1 Microscope should be used and placed in a cool dry non du...

Page 8: ...t again Bulb is broken Replacing bulb Brightness of view field isn t enough or is Uneven Bulb specification doesn t meet the requirement Replacing bulb Brightness isn t adjusted correctly Adjust rotat...

Page 9: ...en correctly Cover glass doesn t meet the requirement Use required thickness cover glass 0 17mm Image observed by two eyes aren t in superposition entirely Interpupilary distance isn t adjusted correc...

Page 10: ...tribution 20 80 Trinocular Head For Infinite System Inclined 30 Rotatable 360 Interpupilary Distance 48 75mm Light Distribution 20 80 Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76...

Reviews: