04/03
347
LB 444
63
8.2
Radiating Interference Detection
Select the function
Radiating interference
(
Parameter
menu, page 19) to suppress inter-
fering radiation (e.g. weld seam testing).
Caution:
When this function is on (Radiating Interference <ON>), the automatic switchover of the
time constant (function
Rapid switchover
) is always off.
When turning this function on, you have to define a
Sigma
value which defines a window
for the measuring signal. If the detector signal suddenly exceeds this threshold,
•
the measurement stops
•
the error message “
Interference Radiation”
appears on the display.
The measured value and the current outputs are held on the last value.
Reset the error message and start the measurement via the
keyboard
by confirming the
error message with <
enter
>; then start the measurement with <
run
> or
externally
via
the digital input.
Note:
Enter n > 5 to rule out false alarms with sufficient statistical safety. The mathematical
correlation shows that the distance from the alarm threshold is dependent upon the re-
spective mean pulse rate I
m
.
For calculation it holds: Sigma =
I s
/
Example:
Count rate I
m
= 300 cps, n = 6
I
S
= I
m
+ n *
m
I
I
S
= 300 + 6 *
300
= 404 cps
Thus, an alarm is signaled as soon as I
S
exceeds the value of 404 cps.
Note:
Due to the dynamic behavior of radiating interfering detection, a quick increase of the
pulse rate due to operative factors (e.g. very fast emptying of the vessel or major short-
term density changes caused by stirrers) may be interpreted as interfering radiation.
For example, opening the active beam channel on the shielding can also cause a quick
increase of the pulse rate. You have to reset the alarm that is then triggered, or better,
do not enable radiating interfering detection at first. Enable radiating interfering detec-
tion only after calibration has been performed.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...