ln I
Density/Concentration
c
a
b
Line a: absorption coefficient correct
Line b: absorption coefficient too small
Line c: absorption coefficient too big
The curve obtained with one-point calibra-
tion always passes through the calibration
point. The gradient of the curve is deter-
mined by the entered absorption coeffi-
cient and may therefore be somewhat in-
accurate.
One-point calibration is also advisable if
only calibration points with very minor
density differences can be recorded, for in
this case minor errors in the lab or during
sampling will significantly affect the curve.
One-point calibration may later be supple-
mented and improved by additional cali-
bration points.
Figure 25:
Influence of absorption coefficients on one-
point calibration
cp
s
a
b
Calibration points
Density in g/cm
3
cp
s
a
b
Calibration points
Density in g/cm
3
Figure 26: One-point calibration with additional calibration points
How to proceed for one-point calibration is described in chapter 6.2.2 Calibration.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...