04/03
347
LB
444
26
4.6
Live Display
Start measurement
A measurement can be started from any menu item by pressing the <
run
> button (ex-
ception: Service menu). The measured values are displayed continuously in the selected
unit for the selected product.
RUN
appears on the display.
With <
+
> and <
-
> you can cycle through the displays below.
Menu sk2: Live Display
Results are displayed in this menu if
RUN
appears on the display; otherwise the values
of the last measurement. Push <
+
> and <
-
> to scroll through the displays below.
Density display
The actual density can be displayed in the defaulted range, e.g. 1.234 g/cm
3
.
I-mean
Shows the average count rate. The entered basic time constant is used as time constant.
I-actual
Shows the actual count rate. The entered basic time constant is used as time constant.
HV auto:
Shows the adjusted high voltage (HV).
Mass Flow
Actual concentration [%] and mass flow [kg/h] are displayed in the defaulted range.
Integr. M=[t]
Shows the integrated mass.
Flow =[m
3
/h]
Shows the actual volume flow.
Stop measurement by pressing <
run
> again.
Select product in the
sk1: Parameter
menu.
4.7
Service Menu
See chapter 10 Service.
4.8
Mass Flow
Unit Mass Flow
For mass flow measurement: select the unit for result output – either kg/h or t/h.
Volume flow at 20 mA
Enter the volume flow in m
3
/h at 20 mA.
Flow
Displays the flow in m
3
/h.
Ext. Mass Counter
Tons per cycle
As soon as the mass flow entered here is reached, the respective relay picks up briefly.
The pick-up time can be set in the Service menu.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...