04/03
347
LB 444
43
If the product is very inhomogeneous, take several samples in quick succession and cal-
culate an average value from their density or concentration values.
Stop measurement by pressing the <
run
> button again.
Accept result with <
enter
>. Cursor jumps to next row.
Enter the density value of the sample(s) in g/cm
3
determined in the lab and confirm with
<
enter
>. Cursor jumps to third row.
Push <
enter
> to confirm current product temperature.
Rates 2. to 10.
must all contain “0”. If “0” is already there, scroll through the rates with
<
more
> until you get to the submenu group
Data input
/
Calculate
. Otherwise, delete
values with <
clear
> and confirm each with <
enter
>.
Push <
sk2
> to select
Calculate
Select calibration mode “
one”
= one-point calibration
Push <
enter
> to confirm the default value for measurement in g/cm
3
or calculate the
absorption coefficients and enter it in
Result a1
.
Calibration starts as soon as you confirm with <
enter
>.
Zero count rate I0
is displayed. Continue with <
more
>.
The entered linear absorption coefficient for this application is displayed. It must not
change during one-point calibration. If it has changed, at least one more value pair is
available.
Correction:
Delete the additional value pair.
Run through steps 15 – 17 once more.
Push <
more
> to skip other coefficients.
If necessary, enter a
factor
for multiplicative correction of the measured values to correct
the gradient of the linear line. This becomes evident only during measurement and is
only required when the absorption coefficient is too small or too big.
If necessary, enter an
Offset
for additive correction of the measured values to offset the
straight line on the Y-axis.
With TC: display of temperature-corrected density values.
Push <
done
> to return to submenu group
Data / Calculate
.
Push <
run
> to start measurement.
For later changes of the current output limit values, calculate calibration curve once
more, since during calculation of the curve the curve is checked for monotony.
Two- and Multi-Point Calibration
The gradient of the calibration curve can be determined accurately. Select the calibration
mode depending on the number of value pairs.
Prerequisite:
Pipeline or container must be filled completely.
Push <
more
> to select menu group
Calibrate
/
Live Display
.
Push <
sk1
> to call the
Calibrate
submenu.
Select product.
Specify if it is a suspension measurement.
Select unit.
Push <
sk1
> to select the
Data input
submenu.
Deselect
Calibr. Data transfer
with “
No”
Read in
1. Rate
: cursor appears in the row “
1. Rate = cps”
.
Push <
run
>. Count rate and temperature (if a temperature sensor is connected) are
read in. Wait until the measuring value has become stable (20 to 50 s). While reading-in
the count rate, the density of the products must not change.
While reading in the count rate, take a sample of the product from the pipeline and de-
termine its density or concentration. If the product is very inhomogeneous, take several
samples in quick succession and calculate an average value from their density or concen-
tration values.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...