BENNING MA 4 measuring adapter
13
6.2.5. R
Insu
4
–
(LN-accessible parts without PE)
Portable test sample
Connection diagram
Circuit diagram
6.2.6. I
PE
– protective conductor current (alt. leak.)
Test procedure (VDE 0701-0702)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the I
PE
measurement on your appliance tester.
As a result, you will obtain the measured I
PE
value of the test
sample.
Application
Portable test sample
Connection diagram
Circuit diagram
6.2.7. I
Cont
– Contact current (alt. leak.)
Test procedure (VDE 0701-0702)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the I
Cont
measurement on your appliance tester.
Contact the test probe(s) of the appliance tester with all
necessary measuring points.
As a result, you will obtain the measured I
Cont
value of the test
sample.
Application
Portable test sample
Connection diagram
Circuit diagram
The contact current is determined for devices of
protection
class II
that raise concerns as to an R
Insu
measurement. Moreo-
ver, it is checked to prove the proper insulation capacity under
mains voltage. This also applies to devices of
protection class I
with accessible conductive parts which are
not
connected to the
protective conductor.