BENNING MA 4 measuring adapter
14
6.2.8. I
Leak
– Device leakage current (alt. leak.)
Test procedure (VDE 0751-1)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the I
Leak
measurement on your appliance tester.
Contact the test probe(s) of the appliance tester with all
necessary measuring points.
As a result, you will obtain the measured I
Leak
value of the test
sample.
Application
Portable test sample
Connection diagram
Circuit diagram
6.2.9. I
PLeak
– Patient leakage current (alt. leak.)
Test procedure (VDE 0751-1)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the I
PLeak
measurement on your appliance tester.
Contact the test probe(s) of the appliance tester with all
necessary measuring points.
As a result, you will obtain the measured I
PLeak
value of the
test sample.
Application
Portable test sample
Connection diagram
Circuit diagram