BENNING MA 4 measuring adapter
10
6.1.9. I
PLeak
–
Patient leakage current
Test procedure (VDE 0751-1)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Place the test sample onto an insulated surface.
Start the
I
PLeak
measurement on your appliance tester.
The test sample is supplied with mains voltage and must
be tested in all switch positions (device functions).
Use the test probe to scan all accessible conductive
parts and applied parts.
As a result, you will obtain the measured
I
PLeak
value of the
test sample.
Connection diagram
Circuit diagram
6.1.10. Functional test
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
The test sample is supplied with mains voltage and must
be tested in all switch positions (device functions).
Please observe inductive/capacitive circuits.
As a result, you will obtain the measured I
PE
value of the test
sample.
Connection diagram
Circuit diagram