BENNING MA 4 measuring adapter
9
6.1.7. I
Cont
–
Contact current
Test procedure (VDE 0701
– 0702 / 0544-4)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the I
Cont
measurement on your appliance tester.
The test sample is supplied with mains voltage and must
be tested in all switch positions (device functions).
Use the test probe to scan all accessible conductive
parts of the test sample that are not connected to the
protective conductor.
As a result, you will obtain the measured I
Cont
value of the test
sample.
Connection diagram
Circuit diagram
6.1.8. I
Leak
–
Device leakage current
Test procedure (VDE 0751-1)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the I
Leak
measurement on your appliance tester.
The test sample is supplied with mains voltage and must
be tested in all switch positions (device functions).
Use the test probe to scan all accessible conductive
parts and applied parts.
As a result, you will obtain the measured I
Leak
value of the test
sample.
Connection diagram
Circuit diagram