BENNING MA 4 measuring adapter
8
6.1.5. R
Insu
4 (LN-accessible parts without PE)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the R
Insu
measurement on your appliance tester.
Use the test probe to scan all accessible conductive
parts of the test sample that are not connected to the
protective conductor.
As a result, you will obtain the measured R
Insu
value of the test
sample.
The testing voltage is applied to the bridged L and N conductors
and measured to parts without PE connection.
Connection diagram
Circuit diagram
6.1.6. I
PE
– Protective conductor current
Test procedure (VDE 0701
– 0702 / 0544-4)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the I
PE
measurement on your appliance tester.
As a result, you will obtain the measured I
PE
value of the test
sample.
The test sample is supplied with mains voltage and must be
tested in all switch positions (device functions).
Please observe inductive/capacitive circuits.
Connection diagram
Circuit diagram