BENNING MA 4 measuring adapter

 

 

 

1.  Application 

The measuring adapter helps you to test portable devices. Due 
to its robust design, it is particularly suited for being used in in-
dustrial environments. The measuring adapter is easy to use and 
thus allows the user to work efficiently and quickly. 

You can test devices and extension cables being provided with 
the following connectors complying with 

IEC 60309-1

CEE connector: 

  16 A,  50 Hz to 60 Hz,  230 V, 

L+N+PE, 6h 

  16 A,  50 Hz to 60 Hz,  380 V to 415 V,  3L+N+PE, 6h 

  32 A,  50 Hz to 60 Hz,  380 V to 415 V,  3L+N+PE, 6h 

To do this, connect the measuring adapter to an appliance 
tester for testing portable devices according to 

DIN VDE 0701-0702, DIN EN 62353 (VDE 0751-1)

 or 

DIN EN 60974-4 (VDE 0544-4)

.  

For  connection  of  the measuring  adapter,  the  appliance  tester 
must be equipped with a shock-proof socket. 
Depending on the type of test, a CEE 16 A or 32 A socket may 
also have to be available. For the CEE 16 A, an optional adapter 
cable (part no. 044163) is required.  

The measuring adapter is suited to be connected to the fol-
lowing appliance testers: 

  BENNING ST 725 

  BENNING ST 750 / ST 750 A / ST 755 / ST 760 

  Devices of other manufacturers 

The measuring adapter can be used for the following tests: 

  Protective conductor resistance (R

PE

  Insulating resistance (R

Insu

  Functional test and phase sequence test of cables 

  Alternative leakage current: 

e. g. protective conductor current (I

PE

), contact current 

(I

Cont

),  

device leakage current (I

Leak

), patient leakage current 

(I

PLeak

  Differential current measuring method 

  I

PE

 

  Leakage current 

 

Direct measurement 

 

Contact current 

 

Patient leakage current 

  Functional test 

  Ua 

  Welding equipment (ST 760) 

  PRCD 

 

The  manufacturers’  specifications  and  normative  re-
strictions  with  regard  to  alternative  leakage  current 
testing and appliance tests must be adhered to. 
To do this, thoroughly read the notes stated in the cor-
responding standard, in the manufacturers’ specifica-
tions of the test sample and in the operating manual of 
the appliance tester.

 

1.1. General functional description  

The MA 4 measures the fault current from the test sample using 
the differential current measuring method. This current is applied 
on the N and PE line of the tester, so that this fault current can 
be measured by testers using the differential current measuring 
method or the direct measuring method with 1 

kΩ in the PE of 

the tester. Deviating measuring resistances in the PE can lead to 
an incorrect current applied. 

2.  Safety notes and precautions 

2.1. Basic information about safety 

The measuring adapter has been designed and tested in compli-
ance 

with 

the 

safety 

requirements 

according 

to 

IEC/ EN 61010-1/ VDE 0411-1

.  

All electrically conductive parts of the housing are connected to 
the protective conductor system (according to 

IEC 61140

pro-

tection class I

).  

The measuring adapter must be used only for testing in environ-
ments 

of 

measuring 

category 

CAT II

 

according 

to 

IEC/ EN 61010-1

Make sure not to connect and test more than one test sample or 
one extension cable at a time. 

The measuring adapter must be operated under supervision 
only. 
In case of proper use according to the specifications, the safety 
of both the user and the device is ensured.  

Read this operating manual and the operating manual of the ap-
pliance tester carefully and thoroughly before using the product. 
Please observe all instructions of the relevant operating manu-
als. 

The measuring adapter must not be used: 

  with the housing being open, 

  in case of visible external damage, 

  in case of damage of the CEE connectors, CEE sockets, 

connecting line or connector, 

  in case of strong excessive strain or if the load limits 

stated in the technical data have been exceeded, 

  if the device has been stored under unfavourable condi-

tions for a longer period of time (high exposure to dust, 
moisture or excessive temperatures). 

Non-observance  might  involve  damaging  or  destruction  of  the 
measuring adapter and/or the connected test sample! 

2.2. Symbols used 

 

Attention! Danger! Please observe documentation! 

 

 

Warning of dangerous voltage! 

 

Note to be observed imperatively 

CAT II 

Device of measuring category II (max. 300 V) 

 

Ground (voltage against ground) 

 

This symbol on the measuring adapter means that 
the measuring adapter complies with the relevant 
EU directives.  

 

The device must not be disposed of via the domes-
tic waste. Further information regarding the WEEE 
mark can be accessed on the Internet by entering 
the search term “WEEE”. 

   

Instructions 

 

 

 

Summary of Contents for 044162

Page 1: ...5116 de Version 1 04 10218133 08 09 2020 BENNING MA 4 CEE measuring adapter 5 pin 3 pin Operating manual Original English version...

Page 2: ......

Page 3: ...Functional test 10 6 2 Testing of three phase devices with alternative leakage current method 11 6 2 1 RPE Protective conductor resistance 11 6 2 2 RInsu 1 LN PE 12 6 2 3 RInsu 2 sec PE 12 6 2 4 RIns...

Page 4: ...nt measuring method This current is applied on the N and PE line of the tester so that this fault current can be measured by testers using the differential current measuring method or the direct measu...

Page 5: ...sed by the protective conductor line of the measuring adapter In case of measuring results near the admissible limiting value measure the protective conductor resistance of the measuring adapter at th...

Page 6: ...BENNING MA 4 measuring adapter 3 4 Operating elements 12 15 14 13...

Page 7: ...ction The testing personnel can use this to detect or limit faults during the cable function test The switch is not used during the functional testing of devices 8 Rotary switch Function This rotary s...

Page 8: ...when 230 V are applied to their test socket For appliance testing the MA 4 must be con nected to the mains via the CEE 32 A cable If such a functional test is carried out no fault current is gen erate...

Page 9: ...ce testing Portable test sample Connect the CEE connector of the test sample to the CEE socket of the measuring adapter Start the RPE measurement on your appliance tester Use the test probe to scan al...

Page 10: ...stance is measured between the PE of the test socket and the secondary side probe Connection diagram Circuit diagram 6 1 4 RInsu 3 LN sec Portable test sample Connect the CEE connector of the test sam...

Page 11: ...le The testing voltage is applied to the bridged L and N conductors and measured to parts without PE connection Connection diagram Circuit diagram 6 1 6 IPE Protective conductor current Test procedure...

Page 12: ...onnected to the protective conductor As a result you will obtain the measured ICont value of the test sample Connection diagram Circuit diagram 6 1 8 ILeak Device leakage current Test procedure VDE 07...

Page 13: ...ons device functions Use the test probe to scan all accessible conductive parts and applied parts As a result you will obtain the measured IPLeak value of the test sample Connection diagram Circuit di...

Page 14: ...Leak device leakage current IPLeak patient leakage current For alternative leakage current measurement the measuring adapter must not be connected to the mains via the CEE 32 connector Single phase de...

Page 15: ...est procedure VDE 0701 0702 Portable test sample Connect the CEE connector of the test sample to the CEE socket of the measuring adapter Start the RInsu measurement on your appliance tester As a resul...

Page 16: ...st procedure VDE 0701 0702 Portable test sample Connect the CEE connector of the test sample to the CEE socket of the measuring adapter Start the ICont measurement on your appliance tester Contact the...

Page 17: ...in the measured ILeak value of the test sample Application Portable test sample Connection diagram Circuit diagram 6 2 9 IPLeak Patient leakage current alt leak Test procedure VDE 0751 1 Portable test...

Page 18: ...r must be set to the position for cable testing 6 3 2 RPE e g CEE 5 pin 32 A Test procedure VDE 0701 0702 Cable test Connect both cable connections to the corresponding CEE sockets of the measuring ad...

Page 19: ...st and phase sequence test of ca bles may only be carried out after the RPE and RInsu tests have been passed For single phase connecting cables or extension ca bles CEE socket 3 pin the rotary switch...

Page 20: ...BENNING MA 4 measuring adapter 17 Circuit diagram...

Page 21: ...dual current protection device RCD used The following PRCDs RCDs can be tested PRCD S PRCD S PRCD K PRCD AC PRCD A PRCD F PRCD B PRCD B PRCD 2 pin PRCD 3 pin Mobile power distributors RCD types A B Fo...

Page 22: ...n with the ST 755 760 Tripping current measurement Tripping time measurement Test key measurement 7 Internal RPE resistance of the measuring adapter RPE measuring adapter Start the RPE measurement on...

Page 23: ...rotection category IP 40 with the case being open IP 67 with the case being closed Dimensions H x W x D mm 170 x 410 x 350 Weight 4 2 kg 9 Maintenance Designation Part no Measuring adapter 044162 The...

Page 24: ...areparts benning de General service requests Phone 49 2871 93 556 E mail servicerequests benning de Returns management Phone 49 2871 93 554 E mail returns benning de Training management Phone 49 2871...

Page 25: ...the technical status at the time of printing Subject to technical changes No liability for printing errors Support helpdesk BENNING helpdesk team Phone 49 2871 93 555 Fax 49 2871 93 417 E mail helpde...

Reviews: