Overview Modbus registers (RCMS150-01 only)
31
RCMS150_D00259_01_M_XXEN/08.2021
Status K1…6
Tab. 6.3: Status K1…6
Error codes
Register 01036
Bit number
Description
0
DC prewarning
1
DC main alarm
2
RMS prewarning
3
RMS main alarm
4
Manual self test
5
Device error
6…31
Reserved
Error
code
Error
Description
Action
3.40
Channel error
Possible error cause:
Temperature too high.
Check if ambient temperature is within the per-
missible range. Check if supply voltage DC 24 V is
within the permissible range. Check if the resid-
ual current through the channel is too high or if
there are high pulses.
If the above points do not apply and the error
occurs frequently, return the device.
6.00
Calibration
error
Calibration data faulty.
Switch the device off and on again. If the error
persists, return the device.
6.50
Production
data faulty
Values outside the limits or
checksum incorrect.
Error is only cleared by switching the device off/
on. If the error persists, return the device.
7.10
Internal
communica-
tion error
Device-internal communication
is disturbed.
Switch the device off and on again. If the error
persists, return the device.
7.62
8.46
Internal
supply voltage
Impermissible deviation
If the error occurs frequently, return the device.