EN
Installation and operating manual
METPOINT
®
BDL compact
55
Home
Settings
Sensor settings
A1
right arrow (2. page)
Gas constant
Preselection of matching
Units
.
The remaining text fields can be labelled and configured in the same manner. For details, see chapter
11.2.2.7 Labelling and
configuring text fields!
Home
Settings
Sensor settings
A1
right arrow (2. page)
If a text field is displayed with red text, the
respective values (e.g. Diameter or Name) have been
changed.
Note:
After confirming with OK, the field labels change back to black and the settings are applied.
Caution:
Reference temperature and reference pressure (factory settings 20°C, 1000 hPa): All volume flow (m³/h) and
consumption (m³) values shown on the display refer to 20°C and 1000 hPa (according to ISO 1217). Alternatively, enter
0°C and 1013 hPa (=standard cubic metre according to DIN 1343) as the reference values. Do not enter the operating
pressure or the operating temperature as the reference values!