APPENDIX 2
IECEx Certification

A2.0   The IECEx Certification Scheme
IECEx  is  a  global  certification  scheme  for
explosion  protected  products  which  aims  to
harmonise international certification standards.
For  additional  information  about  the  IECEx
certification  scheme  and  to  view  the  BEKA
associate certificates, please visit www.iecex.com

A2.1  IECEx Certificate of Conformity
The  BA448CF-F  Fieldbus  Indicator  has  been
issued  with  an  IECEx  Certificate  of  Conformity
number  IECEx  ITS  06.0013X  which  specifies  the
following certification codes and marking:

Ex ia IIC T4 Ga
FISCO Field Device Ex ia IIC T4
Ta = -40

o

C to 70

 o

C

See below for details of safe use in Zone 0.

CAUTION

Installation in Zone 0

When  installed  in  a  Zone  0  potentially
explosive atmosphere, the indicator shall be
installed  such that even in the  event  of  rare
incidents,  an  ignition  source  due  to  impact
or friction between the aluminium enclosure
at the rear of the instrument mounting panel
and iron/steel is excluded.

When  connected  to  an  IECEx  certified  system  the
BA448CF-F may be installed in:

Zone 0

explosive gas air mixture
continuously present.
Note: Special conditions for
safe use apply see above.

Zone 1

explosive gas air mixture
likely to occur in normal
operation.

Zone 2

explosive  gas  air  mixture  not
likely to occur, and if it does will
only exist for a short time.

Be used with gases in groups:

Group   A    propane
Group   B    ethylene
Group   C    hydrogen

Having a temperature classification of:

T1

450oC

T2

300oC

T3

200oC

T4

135oC

At an ambient temperature between –40 and
+70

o

C.

Note: the guaranteed operating temperature range
of the BA448C Fieldbus Indicator is –20 to +70

o

C

The  IECEx  safety  parameters  are  identical  to  the
ATEX parameters  and,  like  the  ATEX  certification,
confirm  that  the  BA448CF-F  complies  with  the
FISCO  Field  Device  requirements  specified  in
IEC60079-27.

The  IECEx  certificate  may  be  downloaded  from
www.beka.co.uk,  www.iecex.com  or  requested
from the BEKA sales office.

A2.2  Installation
The  BA448CF-F  IECEx  and  ATEX  certifications
have  identical  intrinsic  safety  parameters  and
installation  requirements.    The  ATEX  system
design  requirements  described  in  section  4  of  this
manual  may  therefore  be  used  for  IECEx
installations,  but  the  local  code  of  practice  should
always be consulted.

Summary of Contents for BA448CF-F

Page 1: ...Issue 3 23 rd June 2011 BA448CF F FOUNDATION fieldbus Intrinsically safe Panel mounting 8 variable Fieldbus Indicator Issue 3...

Page 2: ...arameters 6 3 5 Enter address display parameters 6 3 6 Enable selected input 6 3 7 Enter Device Address 6 3 8 Enter Variable Identity 6 3 9 Decimal point position 6 3 10 Configuration of bargraph 6 3...

Page 3: ...indicates the protocol but is not shown on the instrument certification label There is an alternative version of the fieldbus display order code BA448CF P for use on PROFIBUS PA networks 1 1 Documenta...

Page 4: ...perature between 40 and 70 o C Note minimum guaranteed operating temperature is 20 C 3 3 Fieldbus connection The BA448CF F Indicator is powered and communicates via the fieldbus which is connected to...

Page 5: ...O applications the BA448CF F Fieldbus Indicator has a higher voltage entity intrinsic safety input parameter allowing connection to a wide range of fieldbus segments The BA448CF F Fieldbus Indicator m...

Page 6: ...that even in the event of rare incidents an ignition source due to impact or friction between the aluminium enclosure at the rear of the instrument mounting panel and iron steel is excluded Fig 2 BA4...

Page 7: ...nditioned as a fieldbus listener see section 6 3 13 to change it to a fieldbus node 6 2 1 Reset the indicator to node factory defaults rESEt The BA448CF F may be reset to the factory defaults that are...

Page 8: ...the P button causes the BA448CF F indicator to search the fieldbus segment to which it is connected for all publishers Fig 5 shows how the indicator displays the hexadecimal Device Address and Variabl...

Page 9: ...vice Address and Variable Identity Parameters can not be changed from within this function See section 6 3 4 Edit Enter addresses and display parameters Contains six sub functions that enable known fi...

Page 10: ...10...

Page 11: ...splayed pressing E will allocate the selected process variable to the selected input and return the display to the Device Address and Variable Identity from which another process variable may be selec...

Page 12: ...function pressing P again will reveal the existing bargraph justification The or button will scroll through the four options when the required justification is displayed pressing E will enter the sele...

Page 13: ...ration menu using the or button To enter the function press P which will show that the instrument is configured as a fieldbus listener to change to a fieldbus node press the or button which will toggl...

Page 14: ...normal operation after a few seconds the model number followed by the firmware version will be displayed prior to entering the display mode 7 2 Fault finding after commissioning ENSURE PLANT SAFETY BE...

Page 15: ...und the liquid crystal display If specified when the instrument is ordered this can be supplied printed with units of measurement and a scale for the horizontal bargraph 8 2 Tag number The BA448CF F c...

Page 16: ...BA448CF F complies with the FISCO Field Device requirements specified in IEC60079 27 The intrinsically safe system shown in Fig 1 of this manual may therefore be used for installations in the USA and...

Page 17: ...17...

Page 18: ...18...

Page 19: ...19...

Page 20: ...20...

Page 21: ...21...

Page 22: ...22...

Page 23: ...23...

Page 24: ...BA448CF F may be installed in Zone 0 explosive gas air mixture continuously present Note Special conditions for safe use apply see above Zone 1 explosive gas air mixture likely to occur in normal oper...