40
When an hourglass appears, open the
clamp while the FeNObreath™ performs
the Quality Check with gas. The device will
automatically draw gas from the Quality
Check Bag.
A reading will appear on-screen; a
successful Quality Check will display an NO
reading within ±10% of the gas
concentration written on the Quality Check
Bag label. For guidance please refer to the
Quality Check conversion chart which can
be found in the Calibration Kit Quick Start
Guide.
If the NO reading is not within ±10%
of the concentration written on the Quality
Check Bag label, your FeNObreath™
requires a calibration. To calibrate your
device with the CaliBag®, follow the
instructions provided.
Ensure both Bag Clamps are closed before
removing the Quality Check Bag and all
adapters from the FeNObreath™.