58
30%.
Electrical fast
transient/bursts
(immunity)
IEC 61000-4-4
±2 kV
±2 kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge (immunity)
IEC 61000-4-5
± 0.5, 1.0 kV L-L
± 0.5, 1.0, 2.0 kV L-E
± 0.5, 1.0 kV L-L
± 0.5, 1.0, 2.0 kV L-E
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
100% 0.01 Seconds
100% 0.02 Seconds
30% 0.5 Seconds
100% 5 Seconds
100% 0.01 Seconds
100% 0.02 Seconds
30% 0.5 Seconds
100% 5 Seconds
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
FeNObreath™ requires continued
operation during power mains
interruptions beyond that provided
by the battery, it is recommended
that the FeNObreath™ is powered
from an uninterruptible power
supply.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical commercial or hospital
environment.
Electromagnetic Immunity
The FeNObreath™ is intended for use in the electromagnetic environment specified below. The customer or the user of
the FeNObreath™ should assure that it is used in such an environment.
Immunity Test
Compliance Level
Electromagnetic Environment –
Guidance
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms (1kHz 80%) 150kHz – 80MHz
6 V rms in ISM and amateur radio
bands
3 V/m (1kHz 80%) 80MHz – 2.7GHz
The FeNObreath™ is suitable for the
electromagnetic environment of typical
commercial or hospital settings.
FeNObreath™ has also been tested for immunity to RF wireless communications equipment as be.