19
Press the red ‘x’ to confirm.
The profile will be deleted and the profile screen
will be displayed.
Maintenance
FeNObreath™ - FeNO testing without limits
The FeNObreath™ has been validated for up to 29,000 tests when used as instructed and properly
maintained and serviced. The number of tests can be periodically checked within the settings of the
device; when 29,000 tests are reached a service is recommended. Contact MGC Diagnostics
Corporation or the local service center.
The Health Care Professional can check how many breath tests have been performed on the device
using the Information Screen, Test No as seen below.
Routine maintenance
1.
Mouthpieces are to be replaced after every patient.
WARNING: The mouthpieces are single patient use only and can be used for a maximum of
3 tests. Further re-use could cause incorrect readings and could increase the risk of cross
infection. The mouthpiece should be disposed of after use, in accordance with local waste
disposal guidance.
2.
Hands should be washed regularly in accordance with infection control practice.