Possible Cause
Recommended Action
The device battery is
flat/empty
Place the monitor in the
charging dock and plug
into a power source.
Alternatively, plug the
device directly into a
power source. See
‘Installation and Setup’
section of this manual.
Suspicion that the device is giving
erroneous/incorrect readings
.
This image is an example only, and is not necessarily an example of an erroneous reading.
Possible Cause
Recommended Action
The device may be out
of specification.
If the user suspects the device
is giving erroneous readings,
stop using it and contact MGC
Diagnostics Corporation or the
local service center.
The device may have
been exposed to high
levels of volatile organic
compounds (VOC’s) for
example from cleaning
agents.
Allow the device to rest for up
to 24 hours in a VOC free
environment.
The device may be
showing testing in
demo mode.
Ensure the breath test mode is
being selected from the home
screen.