3-12
Operation
The default setting for both acceleration and deceleration is
maximum (MAX) for all rotors except the JS-24.38 and the JS-24.15.
SLOW is the default setting for these rotors; TIMED settings can also
be used.
Acceleration
Action
Result
1. Press [
A/D
]. The
Accel:
field flashes. MAX, SLOW, and TIME soft keys appear.
2. Press the appropriate
soft key.
If you choose MAX or SLOW, your selection appears in the
Accel:
field. If you choose TIME, enter an acceleration time from 1 to
10 minutes and press [
A/D
] again or accept the 10-minute default
setting.
Deceleration
Action
Result
1. Press [
A/D
] twice (if the
cursor is in the
Accel:
field,
press [
A/D
] once).
The
Decel:
field flashes. MAX, SLOW, TIME, and OFF soft keys
appear.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Rotor : JA-25.50
Speed : 25 000 RPM = x G
Time : 00:20
Temp : 20°C Max Temp : 24°C Shutdown? No
MAX
SLOW
TIME
Accel : 5 Decel :
Enter 1 to 10 minutes from 0 to 500 rpm