![Beckman Coulter Avanti J-25 Instruction Manual Download Page 16](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-25/avanti-j-25_instruction-manual_2719702016.webp)
1-3
Specifications and Preinstallation Requirements
AVAILABLE ROTORS
Refer to the applicable rotor manual for complete rotor descriptions.
a
Maximum speeds are based on a solution density of 1.2 g/mL with the following exceptions: the JA-18.1 rotor is rated for a
density of 1.4 g/mL; the JCF-Z rotor is rated for a density of 1.45 g/mL; and the JE-6B rotor is rated for a density of 3 g/mL.
b
Maximum speed for the JA-30.50 Ti rotor is 25 000 rpm in the Avanti J-25I centrifuge, and 30 000 rpm in the Avanti J-30I centrifuge.
c
Maximum speed for the JS-24.38 and JS-24.15 rotors is 10 000 rpm in the Avanti J-25I centrifuge, and 24 000 rpm in the
Avanti J-30I centrifuge.
Rotor Profile and Description
Max
RPM
a
Max RCF
(
×
g
)
Max
Capacity
Rotor
Manual
Number
JA-30.50 Ti Fixed Angle, 34° (8 place)
r
max
= 108 mm
25 000
b
75 600
8
×
50 mL
J-TB-070
JA-25.50 Fixed Angle, 34° (8 place)
r
max
= 108 mm
25 000
75 600
8
×
50 mL
J-TB-056
JA-25.15 Fixed Angle, 25° (24 place)
r
max
= 106 mm
(outer row)
r
max
= 86 mm
(inner row)
25 000
74 200
(outer row)
60 200
(inner row)
24
×
15 mL
J-TB-057
JA-21 Fixed Angle, 40° (18 place)
r
max
= 102 mm
21 000
50 400
18
×
10 mL
J-TB-002
JA-20.1 Fixed Angle, 23° (32 place)
r
max
= 115 mm
(outer row)
r
max
= 98 mm
(inner row)
20 000
51 500
(outer row)
43 900
(inner row)
32
×
15 mL
J-TB-022
—Continued