1-11
Specifications and Preinstallation Requirements
• Allow a 7.7-cm (3-in.) clearance on each side of the instrument
and a 16-cm (6.25-in.) clearance behind the instrument for air
circulation. The centrifuge must have adequate air ventilation to
ensure compliance to local requirements for vapors produced
during centrifugation operation.
Position the centrifuge so that the air diverter, shown in Figure 1-4,
touches the wall behind the centrifuge. Place the power cord to
one side of the air diverter. To avoid damaging the power cord
when installing or moving the centrifuge, be sure to move the cord
out of the way before pushing the centrifuge against the wall.
Figure 1-4. Rear View and Dimensions
cm
in.
6.0 ft
1.8 m
6.25
16
Air Exit
Depth:
Air Intake
Air Diverter
Power Cord
71
28
116
45.5
141
55.5
86
34
86
34