3-7
Operation
Action
Result
2. Press the RCF soft key.
The RPM = - -
×
G
field flashes.
3. Enter the required rcf on the
keypad and press [
ENTER
].
The rcf and the corresponding rpm appear in the
Speed:
field. The
RCF (
×
g
) indicator illuminates on the actual value display, indicating
that rcf (
×
g
) is being displayed during the run. When [
ENTER
] and
[
START
] are pressed, the centrifuge runs at the calculated rpm.
If you enter an rcf value that is unachievable for the installed rotor,
a message will appear telling you the rotor’s achievable rcf range.
ENTERING RUN TIME
Run time can be entered in hours and minutes or in
ω
2
t
(accumulated
centrifugal effect). Hold mode is used for continuous operation.
Entering a Timed Run
Action
Result
1. Press [
TIME
].
HH:MM, W2T, and HOLD soft keys appear.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
ROTOR ID
SPEED
RPM/RCF
TIME
HR:MIN
°C
TEMP
Rotor : JA-25.50
Speed : - - RPM = 75 600 x G
Time : 00:00
Temp : 25°C Max Temp : 29°C
Accel : MAX Decel :
RPM
RCF