1-5
Specifications and Preinstallation Requirements
a
Maximum speeds are based on a solution density of 1.2 g/mL with the following exceptions: the JA-18.1 rotor is rated for a density of 1.4 g/mL;
the JCF-Z rotor is rated for a density of 1.45 g/mL; and the JE-6B rotor is rated for a density of 3 g/mL.
b
Maximum speed for the JA-30.50 Ti rotor is 25 000 rpm in the Avanti J-2 I centrifuge, and 30 000 rpm in the Avanti J-30I centrifuge.
c
Maximum speed for the JS-24.38 and JS-24.15 rotors is 10 000 rpm in the Avanti J-25I centrifuge, and 24 000 rpm in the Avanti J-30I centrifuge.
Rotor Profile and Description
Max
RPM
a
Max
RCF
(
×
g
)
Max
Capacity
Rotor
Manual
Number
JA-12 Fixed Angle, 35° (12 place)
r
max
= 144 mm
12 000
23 200
12
×
50 mL
J-TB-051
JA-10 Fixed Angle, 25° (6 place)
r
max
= 158 mm
10 000
17 700
6
×
500 mL
J-TB-006
JLA-10.500 Fixed Angle, 20° (6 place)
r
max
= 166 mm
10 000
18 500
6
×
500 mL
J-TB-048
F10BCI-6x500y Fixed Angle, 23° (6 place)
r
max
= 158 mm
10 000
17 696
6
×
500 mL
—
JLA-9.1000 Fixed Angle, 20° (4 place)
r
max
= 185 mm
9 000
16 800
4
×
1000 mL
J-TB-073
JS-24.38 Swinging Bucket (6 place)
r
max
= 161.0 mm
10 000
c
18 000
6
×
38.5 mL
J-TB-058
—Continued