3-8
Operation
Action
Result
2. Use they keypad to enter
the required time (up to
99 hours, 59 minutes) and
press [
ENTER
].
Entered time appears in the
Time:
field.
3. Press the next function key or
press [
ENTER
], then [
START
].
The run begins, and the time remaining begins counting down to 0 in
the
TIME
display. If you enter a number over 59 minutes, the centri-
fuge automatically recalculates the time in hours and minutes after
[
ENTER
] or any function key is pressed.
Deceleration begins when the
TIME
display has counted down to 0.
During deceleration, “End” flashes in the
TIME
display and the red
indicator light on the [
STOP
] key flashes.
Entering an
ω
2
t Run
Action
Result
1. Press [
TIME
].
HH:MM, W2T, and HOLD soft keys appear.
2. Press the W2T soft key to
select the
ω
2
t
mode.
The
ω
2
t
indicators (EXP 00 and rad2/sec) light in the
TIME
display.
3. Use the keypad to enter an
ω
2
t
value.
The entry appears in the
Time:
field. The decimal point and E
appear automatically as you type. If you enter a value out of the
achievable
ω
2
t
range for that rotor, a message prompts you to enter
a different value.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Rotor : JA-25.50
Speed : 25,000 RPM = x G
Time : 00:25
Temp : 25°C Max Temp : 29°C
Accel : MAX Decel : MAX
HH:MM
W2T
HOLD