HTS Customer Care Kits User’s Guide
Replacing Wash Station Components
1
Perform the monthly cleaning procedure to decontaminate the sample
injection port/wash station. For instructions, see your
BD
High
Throughput Sampler User’s Guide
.
2
Turn off the cytometer and HTS if needed.
3
Remove the HTS according to procedure in Chapter 2.
4
Move the probe arm up and toward the front of the HTS.
5
Unscrew and remove the sample injection tubing
from the top of the
secondary pump.
Injection port/wash station
Injection port tubing
Sampler coupler
Sealing washer