32
HTS Customer Care Kits User’s Guide
Required Kit Components
Pump Syringe Replacement Procedures
•
If you are replacing the primary pump,
use both procedures in this chapter.
•
If you are replacing the secondary pump, remove the HTS (see Chapter 2),
then use the Replacing the Syringe procedures.
Removing Liquid from the Syringe (For Primary Pump Syringe)
1
Verify that the cytometer and BD FACSDiva software are running.
2
From the Acquisition Dashboard select
RUN PLATE
.
3
Remove the liquid from the syringe:
a
Disconnect the sheath connector from the back of the HTS.
b
Disconnect one end of the sheath filter.
c
Select
HTS > Prime.
d
Repeat until the syringe barrel is empty.
e
Exit BD FACSDiva software.
4
Turn off the cytometer.
Item number
Part number
Description
Qty
6
344912
Syringe, 500 µL, with black plunger seal
1
7
347306
Lubricant, O-ring, Super O-Lube, ½-
ounce tube
1