MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 90 of 144
10.6. Error Table
The MX08 PN gateway instruments have been designed as very reliable and virtually error
free instruments. However if an error occurs do not attempt to repair the equipment before
you understand what caused the error. Note the problems you have with your instrument
and the error messages shown by the LEDs located on the front panel. Then try to solve
the problem according to the error table given below.
Error Code
Description
Actions to take / Possible cause
Pwr
Link
Err
System Error
− Re-energize the instrument. If seen again, change
the board.
Configuration
Error
− Re-address the instruments on N-Bus (Refer to
section 3.3 or Section 3.9 N-Bus addressing).
No Instrument
Found
− Install MX08 AD or MX08 IO instruments to the
system.
− Check the N-BUS connector whether they are
installed on the rail.
Blank Light
Flash Blank for 0.3 second
Table 9.2 - Error table