MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 74 of 144
MX08
AD[0]
2
nd
Dword
(R)
By default, Indicated Weight value is represented.
To represent other weight or calibration status, refer to next Dword.
3
rd
Dword
(R)
Not in use
Error Table of
Digitizer[0]
Weight
or
Count
Zero
Rang.
Gross
or Net
Motion
Detec
Response of Read Selected
Data
Cmd
Flag
Bit
Number
3
rd
, 5
th
, 7
th
, 9
th
, 11
th
, 13
th
, 15
th
, 17
th
Dword Description
D31
…
D16
Not in use
D15
D14
D13
D12
0111
Instrument does not found
Error Codes of Digitizer
0110
Low/High voltage detection error
0101
In programming mode
0100
System Error
0011
ADC under
0010
ADC over
0001
ADC out
0000
No error found
D11 …
D10
Not in use
D9
0
Weighing / Force Mode
Operation Mode
1
Count Mode
D8
0
Out of Zero Range
Zero Range
1
In Zero Range
D7
0
Gross
Indication
1
Net
D6
0
Stable
Motion Dedection
1
Dynamic
D5
D4
D3
D2
D1
00000
Indicated weight
Response of 2
nd
Dword
description
00001
Gross weight
00010
Tare weight
00011
Indicated weight (floating point type)
00101
Tare weight (floating point type)
10000
Calibration Status (Refer to below table)
D0
Toggles
The command is applied successfully
CMD Flag