MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 42 of 144
Average Load cell Output:
Enter load cell output sensitivity in mV/V. If the weighing
system has more than one load cell, calculate the mean value of load cells outputs mV/V
indicated on the certificates of the load cells.
Example: If load cell outputs are LC1: 2.0010, LC2: 1.9998, LC3:1.9986 and LC4:2.0002,
the mean value will be
LC output = ( 2.0010 + 1.9998 + 1.9986 + 2.0002 ) ÷ 4 = 1.9999 mV/V.
Estimated Dead Load:
Enter the estimated dead load value of the weighing system in kg.
You may perform zero adjustment in convenient time for exact dead load compensation.
After pressing
Save eCal Data to Digitizer
, these data transferred to the instrument and
eCal is finalized.
If the scale is empty and you want to make automatic zero adjustment instead of entering
estimated dead load, press
eCal with Zero Adjustment
button for starting zero calibration.
The display will show [WAIT] message during zero adjustment for approximately 5
seconds. In this while the scale must be unloaded and stable. eCal calibration coefficients
are saved automatically.
eCal Wizard:
eCal Wizard helps the user to calculate Total Load Cell capacity and Average Load cell
Output values via entering each loadcell capacity and mV/Vvalues.
5.4. MX08 AD Status
xFace status tab provides you information of count, gross, tare, net, indication, in zero
range, error, serial number, firmware revisions of MX08 AD which have the selected
address. Taring, Zeroing and Clear keys are located in this tab.