Configurable Options
Item
Descriptor
Options
Factory
Setting
Maximum Flow Rate
MAX RATE
1–1999 mL/hr
1999
mL/hr
Maximum VTBI
MAX VTBI
1–9999 mL
9999 mL
Flow Check Display
FLOW
CHECK
1 = OFF
(The
CLR
&
TOT VOL/STATUS
keys must be pressed to view
the Flow Check display)
2 = ON
(Flow Check display visible
whenever pump is running)
OFF
Baud Rate
(The rate of information
exchange between the
pump and the computer
when in Computer Con-
trol)
BAUD RATE
1 = 300
2 = 1200
3 = 2400
4 = 4800
5 = 9600
9600
Computer Control
(Selects the type of com-
puter control available at
power-up)
COMP CONT
1 = DISABLED
(Computer Control not avail-
able)
2 = OFF W ALRM (the pump
drops out of Computer Control
when an alarm occurs)
3 = ON W ALRM (pump re-
mains in Computer Control
during an alarm
DIS-
ABLED
Hospital Area Designator
(displayed for 3 seconds
after self-test. Custom
designators can be cre-
ated using the optional
Pump Configuration Soft-
ware.)
HAD
0 = No message
1 = NICU
2 = PICU
3 = MED/SURGICAL
4 = TRAUMA/BURN UNIT
5 = OPER ROOM
6 = CARDIAC/ICU
7 = SURGICAL/ICU
8 = ICU
9 = ONCOLOGY
No mes-
sage
42
Rev.B