The OPTIONS Key
The
OPTIONS
key is used to place the pump into Computer
Control mode and to set a Programmed Delivery Profile
(PDP). The
OPTIONS
key is active only when set in the
pump's configuration. Pressing the
OPTIONS
key selects
Computer Control and/or PDP. The PDP has two methods
of programming. The manual Program Delivery method al-
lows you to manually program an infusion with up to 10 dif-
ferent steps. The Ramp PDP method allows you to program
a ramped infusion by entering volume and time parameters.
The availability of the PDP programming option is depend-
ent upon the pump's configuration.
Note:
If the word
OPTIONS
is not visible when the pump is
powered on, you cannot use computer control or PDP.
Consult qualified hospital biomedical personnel re-
garding the appropriateness of the pump's settings.
Using the Manual Programmed Delivery Profile
The manual Programmed Delivery Profile (PDP) allows you
to program an infusion with up to 10 different steps. The
pump automatically changes to the next programmed rate
once the VTBI of a step has been delivered. To use the 10
step manual Program Delivery Profile, proceed as follows:
1.
Verify that the
OPTIONS
key is lit.
a.
If
OPTIONS
is not illuminated, your
pump is not configured for PDP.
b.
Consult with authorized hospital
personnel if you have questions re-
garding the appropriateness of your
pump's configuration settings.
21
Rev.B