Message
Flow
Status
Condition
SEC
COMPLETE
No
change
The SEC COMPLETE alert option is enabled and the pump
has completed its secondary program and has switched to
primary. Press any enabled key to exit from this alert.
SEC
PROGRAM
No
change
Secondary program data is being entered while pump is
running. Pump will remain in alert condition until you
press
SEC START
or
STOP
.
SEC RATE=0
No flow
A secondary flow rate of zero has been entered. The pump
will remain in this alert condition until you enter a non-
zero secondary flow rate and press
SEC START
, or start the
primary program.
SEC VTBI=0
No flow
A secondary VTBI of zero has been entered. The pump will
remain in this alert condition until you enter a non-zero
secondary VTBI and press
SEC START
, or start the pri-
mary program.
FLOW RATE
No flow
During Volume-Time Programming, if a flow rate outside
the pump's capabilities is calculated, the message
Hi
or
Lo
will be displayed in the rate display. If an attempt is made
to start the pump with either of these messages displayed, a
FLOW RATE alert will be triggered. To exit this alert condi-
tion, reprogram the pump for a rate within the range set in
the configuration and press the appropriate
START
key.
CHECK VTBI
No flow
A VTBI outside the acceptable range has been entered. To
exit this alert condition, enter a VTBI within the range set
in the configuration and press the appropriate
START
key.
PGM DELIV
ENTER PGM
No flow
The pump is waiting for manual Programmed Delivery Pro-
file values to be entered. Either enter the values (as de-
scribed on p. 21), or press
OPTIONS
to leave PDP mode.
PGM DELIV
CLEAR ALL?
No flow
The pump is asking for confirmation that all steps of the ex-
isting Programmed Delivery Profile should be erased. Do
one of the following: press
CLR
to erase the profile; press
NEXT
to begin reviewing the rest of the profile; press
PRI
RATE
or
PRI VTBI
to modify this step; or press
OPTIONS
to
leave PDP mode.
34
Rev.B