baxter Flo-Gard 6201 Operator'S Manual Download Page 26

2.

Place the pump into Pro-
gram Delivery Profile
mode by pressing the

OP-

TIONS

key.

Press the key until the message

PGM

DELIV

appears and the

NEXT

legend over

the

TOT VOL/STATUS

key is illuminated.

The pump is now in PDP mode for an in-
fusion with up to 10 different steps. If

RAMP PDP

appears, the pump is config-

ured for automatic ramped delivery.
Consult with authorized hospital per-
sonnel if you have any questions regard-
ing the appropriateness of your pump's
configuration.

3.

Review previous Pro-
gram Delivery steps.

a.

If a previously set Program Delivery
Profile is stored in the pump's mem-
ory,

STEP 1

appears in the message

display and the rate and VTBI of the
first programmed step appear in the
main display.

b.

If you wish to clear the stored Pro-
grammed Delivery Profile, press the

CLR

key. The message

CLEAR ALL?

appears. Press the

CLR

key again to

clear all Program Delivery rates and
VTBIs. The

CLEAR ALL?

message

disappears.

c.

If you wish to use the stored Pro-
grammed Delivery Profile, continue
pressing the

NEXT

key to scroll

through and review all steps of the
Programmed Delivery Profile, until
the first step is again displayed. Re-
view each step to ensure that the
Program Delivery rates and VTBIs
are appropriate.

22

Rev.B

Summary of Contents for Flo-Gard 6201

Page 1: ...Operator s Manual Flo Gard 6201 Volumetric Infusion Pump...

Page 2: ......

Page 3: ...le the Pump Is Running 20 The OPTIONS Key 21 Using the Manual Programmed Delivery Profile 21 Using the Ramp Programmed Delivery Profile 24 Computer Control 29 Computer Monitoring 30 Locking the Front...

Page 4: ...19 01 602 Original issue 4 92 Rev A 6 92 Rev B 4 93 Rev C 11 94 Rev D 3 99 Page Revision Front cover 0 i B ii D 1 B 2 11 0 12 B 13 14 0 15 38 B 39 C 40 46 B Back cover D A zero in the Revision column...

Page 5: ...istration Safety clamp automatically occludes the tubing when the pump door is opened Occlusion sensors detect both upstream and down stream restrictions Sensitivity of the downstream occlusion sensor...

Page 6: ...over to the primary program upon completion Volume Time programming automatically calcu lates flow rate Automatic self test routine checks for proper func tion before use Five hour memory retains inf...

Page 7: ...ered when pressed The key is also used to advance to the next step when programming a delivery profile See NEXT legend item 4 above 6 CLEAR TOT VOL key Resets the total volume delivered to zero when t...

Page 8: ...ART 7 8 9 4 5 6 1 2 3 0 CLR ALARM PRI VTBI PRI RATE ON OFF CHARGE STOP SEC RATE SEC VTBI SEC START OPTIONS NEXT COMPUTER CONTROL MONITOR P U S H VOLUMETRIC INFUSION PUMP Flo Gard 6201 1 2 3 4 5 6 7 8...

Page 9: ...mming values and decimal point are entered with these keys 21 PRI VTBI key Allows programming of the primary VTBI 22 PRI RATE key Allows programming of the primary infusion rate 23 PRI START key Start...

Page 10: ...Air sensor Detects air bubbles in the tubing The sensi tivity is adjustable via the pump s configura tion 5 Safety clamp Prevents accidental fluid flow by automati cally occluding the tubing whenever...

Page 11: ...7 Figure 2 Pump Head Features 1 2 3 4 5 6 7...

Page 12: ...only 5 VOLUME knob Adjusts volume of alert and alarm tones The alarm and alert tones cannot be turned com pletely off 6 PANEL LOCK switch Disables all front panel keys except BACK LIGHT and TOT VOL S...

Page 13: ...9 PANEL LOCK VOLUME COMMUNICATIONS PORT 1 2 PUSH 3 4 5 6 7 8 9 Figure 3 Rear of Pump...

Page 14: ...clinical applica tions Use only with standard Baxter soft tubing admini stration sets with an s designation Read and understand this manual before using the Flo Gard 6201 pump When infusing fluid thro...

Page 15: ...ural delivery It is strongly recommended that pumps used for epidural drug delivery be clearly differentiated from pumps used for other routes of administration WARNING Epidural administration of drug...

Page 16: ...serted Proceed to step c Note During an emergency in the alert mode the pump will operate without inserting the slide clamp An alert tone will sound until the slide clamp is inserted or the SILENCE k...

Page 17: ...pump door and removing set 8 Attach set to I V access site 9 Turn pump on by press ing the ON OFF CHARGE key Verify that the pump performs the fol lowing self test a All segments of both displays app...

Page 18: ...ty clamp Close the door and open the regulating clamp 10 Set VOLUME knob on the rear of the pump to the desired level Starting a Primary Infusion 1 Set primary flow rate as follows a Press PRI RATE b...

Page 19: ...NG LED lights and a moving bar appears next to the flow rate setting b If the alarm software versions later than 1 08 or the alert tone sounds with the message INSERT SLIDE CLAMP when you press PRI ST...

Page 20: ...lert tone and switches to a KVO Keep Vein Open rate of 5 mL hr or the current rate setting whichever is lower 8 Remove set from pump as follows a Press the STOP key b Close administration set regulati...

Page 21: ...ove the pump 3 Lower primary container with the hanger accom panying the secondary set 4 Set secondary flow rate as follows a Press SEC RATE key The right hand side of the main display now shows secon...

Page 22: ...program When this happens if the primary flow rate is set above 999 mL hr be sure to close the secondary set on off clamp Using Volume Time Programming Volume Time Programming allows you to select a V...

Page 23: ...e PRI RATE or SEC RATE key a The pump calculates the flow rate required to deliver the desired VTBI in the specified time period b If the calculated flow rate is higher than the pump s capabilities th...

Page 24: ...n ning NEW RATE appears in the message dis play and an alert tone sounds periodi cally 2 Enter the new flow rate on the keyboard a If the primary rate is changed above 999 mL hr ensure that the secon...

Page 25: ...uration Note If the word OPTIONS is not visible when the pump is powered on you cannot use computer control or PDP Consult qualified hospital biomedical personnel re garding the appropriateness of the...

Page 26: ...ps a If a previously set Program Delivery Profile is stored in the pump s mem ory STEP 1 appears in the message display and the rate and VTBI of the first programmed step appear in the main display b...

Page 27: ...rt to enter a PDP 4 Modify previous Pro grammed Delivery steps a Press the NEXT key to reach the Pro gram Delivery step you wish to modify b Press the key for the infusion pa rameter you wish to modif...

Page 28: ...e same manner After enter ing settings for the last step press NEXT again until the first step is dis played d To begin the Program Delivery Pro file press PRI START The pump will begin delivering flu...

Page 29: ...ing this phase the pump delivers fluid at 5mL hr KVO rate Run Time This represents the total time of the Ramp PDP excluding the delay time Ramp Up Time This represents the time from the start of the t...

Page 30: ...d the NEXT legend over the TOT VOL STATUS key is illuminated The pump is now in the PDP mode If PGM DELIV appears the pump is configured for manual Program De livery Consult with authorized hospital p...

Page 31: ...to modify b Press the CLR key to erase the pre vious parameter c Enter the new value on the numeri cal keyboard d Press the NEXT key to lock in the pro grammed value and advance to the next parameter...

Page 32: ...e NEXT key to lock in the pro grammed value and advance to RUN TIME h Press the CLR key to allow the entry of a new value i Enter the value on the numerical keyboard j Press the NEXT key to lock in th...

Page 33: ...PDP appears in the message display if the pump is turned on within one hour of being turned off b Press PRI START to resume the Ramp PDP and the pump begins delivering fluid The pump will switch to t...

Page 34: ...ont Panel The front panel can be locked during pumping to prevent tampering It can be unlocked at any time The TOT VOL STATUS and BACK LIGHT keys are not affected by the lock out This allows routine i...

Page 35: ...CHECK display is a dashed line that stretches from left to right as the resistance to flow increases The left side of the display is labeled norm and corresponds to the pressure when the infusion was...

Page 36: ...disonite Edison Chemical Co TOR or Hi Tor Plus Huntington Labs Bafix Hysan Corp A solution of 10 bleach and water Soapy water Isopropyl alcohol up to 95 Note Do not clean disinfect or sterilize any pa...

Page 37: ...s not loaded into the slide clamp slot Open the pump door and insert the slide clamp into the slot Close the door NEW RATE No change until proce dure is com plete Flow rate is being changed while pump...

Page 38: ...pabilities is calculated the message Hi or Lo will be displayed in the rate display If an attempt is made to start the pump with either of these messages displayed a FLOW RATE alert will be triggered...

Page 39: ...e the Ramp PDP where it was interrupted Do one of the following press PRI START to start fluid delivery or press STOP to cancel the alert COM TIMEOUT No change There has been no communication between...

Page 40: ...prevent the pump from restarting automatically press any enabled key such as SILENCE during the alarm INSERT SLIDE CLAMP Note This alarm option is available only on pumps running software versions lat...

Page 41: ...e plies The condition has been caused by the computer not the pump To clear the alarm press OPTIONS to return the pump to computer control or press SILENCE or any pro gramming key to control the pump...

Page 42: ...0 Hz Power Cord 2 9 m 9 5 ft long with Hospital Grade plug Fuse 0 5 A 250V SB 5 2 mm 13 64 in 20 mm 25 32 in Leakage Current Less than 50 microamps using UL 544 specified test methods Weight Approxima...

Page 43: ...rsonnel Delay Time 00 00 08 00 hours in minute increments Run Time 01 00 24 00 hours in minute increments Ramp Up 00 00 to 45 of the Run Time in minute in crements Ramp Down 00 00 to the Run Time minu...

Page 44: ...or SEC START key To view each subsequent option press the NEXT or SEC START key again c To stop viewing the configuration again press TIME and TOT VOL STATUS simultaneously Note If there is any quest...

Page 45: ...e self test when the pump is turned on AUDIB SWI 1 OFF 2 ON OFF Number of Automatic Re starts The number of times the pump will restart auto matically after a down stream occlusion occurs and is then...

Page 46: ...trol BAUD RATE 1 300 2 1200 3 2400 4 4800 5 9600 9600 Computer Control Selects the type of com puter control available at power up COMP CONT 1 DISABLED Computer Control not avail able 2 OFF W ALRM th...

Page 47: ...lamp Loading Op tion Note The ALARM op tion is available only on pumps running software versions later than 1 08 INS CLAMP 1 OFF 2 ON for software versions earlier than 1 09 or ALERT for software vers...

Page 48: ...saved indefinitely until changed from the front panel 4 PERM Profile is saved in permanent memory If a new PDP is entered the pump will revert to the PDP in perma nent memory once the cycle has been c...

Page 49: ...y to the original purchaser and is not assignable or transferable and shall not apply to auxil iary equipment or disposable accessories THERE ARE NO OTHER WARRANTIES INCLUDING ANY IMPLIED WARRANTY AND...

Page 50: ...tainer or in another Baxter ap proved container that will provide adequate protection dur ing shipment To ensure prompt return a Baxter Product Service representative must be notified before shipping...

Page 51: ......

Page 52: ...Baxter Healthcare Corporation Deerfield IL 60015 USA Copyright 1994 1999 Baxter Healthcare Corporation All rights reserved 07 19 D1 602...

Reviews: