Note:
If any of the entered parameters (Total VTBI, Run
Time, Delay Time, Ramp Up and Ramp Down) are
outside the pump's capabilities, the message
Hi
or
Lo
appears as appropriate on the main display.
1.
Verify that the
OPTIONS
key is lit.
a.
If
OPTIONS
is not illuminated, your
pump is not configured for PDP.
b.
Consult with authorized hospital
personnel if you have questions re-
garding the appropriateness of your
pump's configurations settings.
2.
Place the pump into
Ramp PDP by pressing
the
OPTIONS
key.
a.
Press the key until the message
RAMP PDP
appears and the
NEXT
legend over the
TOT VOL/STATUS
key
is illuminated. The pump is now in
the PDP mode. If
PGM DELIV
appears, the pump is
configured for manual Program De-
livery. Consult with authorized
hospital personnel if you have any
questions regarding the appropriate-
ness of your pump's configuration.
3.
Review previous Ramp
PDP parameters.
a.
If a previous Ramp PDP is stored in
the pump's memory,
TOTAL VTBI
appears in the message display and
a volume, in mL, appears in the
main display.
b.
If you wish to use the stored Ramp
PDP parameters, press the
NEXT
key
to scroll and review through all pa-
rameters of the Ramp PDP, until the
TOTAL VTBI
message is again dis-
played. Review each parameter to
ensure that the Ramp PDP parame-
ters are appropriate.
26
Rev.B