b.
The
PUMPING
LED goes out and
STOPPED
appears in the message
display.
c.
If the pump is not restarted within
two minutes, an audible alert
sounds.
6.
To restart an infusion,
press the
PRI START
key.
7.
When the infusion com-
pletes:
When the pump has delivered the se-
lected volume, it sounds an alert tone
and switches to a KVO (Keep Vein
Open) rate of 5 mL/hr or the current
rate setting, whichever is lower.
8.
Remove set from pump
as follows:
a.
Press the
STOP
key.
b.
Close administration set regulating
clamp.
c.
Open pump door.
d.
If the slide clamp loading option is
enabled, press the slide clamp fully
into its slot.
e.
Press the
SAFETY CLAMP
to the open
position.
f.
Check that no fluid is flowing in set,
then remove set from pump.
g.
To turn the pump off, press the
ON-
OFF/CHARGE
key.
16
Rev.B