6.
Close pump door.
If resistance is felt when closing the
door, check for misloaded I.V. set.
7.
Open set regulating
clamp completely.
a.
Verify that no drops are falling in
drip chamber. If flow is observed,
close regulating clamp, recheck I.V.
set loading and verify that the
proper administration set is being
used.
b.
If flow is again observed, do not use
the pump. Have it inspected by serv-
ice personnel.
Note:
Always
close
the
administration
set
regulating
clamp(s) before opening pump door and removing set.
8.
Attach set to I.V. access
site.
9.
Turn pump on by press-
ing the
ON/OFF CHARGE
key. Verify that the
pump performs the fol-
lowing self-test:
a.
All segments of both displays appear
momentarily.
b.
Three separate audible tones sound.
c.
If a Hospital Area Designator (HAD)
has been programmed into the
pump, it appears for 3 seconds.
d.
The occlusion detection level is mo-
mentarily displayed in the message
display (
LEVEL 1, 2,
or
3
), followed
by
AUDIBLE SWITCHOVER
if the
Audible Switchover option is en-
abled.
e.
If Auto Restart and Flow Check are
both enabled, the message
AUTO
RESTART
appears for one second
following the occlusion detection
level display.
13