Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 47
8.
Close the pump door.
9.
Connect the end of the outlet tube to the tube holder on the vial rack.
Connecting the outlet tube
Preparing the New Ingredients
WARNING
The compounder is not for use with non-sterile containers.
Source containers that can be used with the compounder are:
•
Large-volume, vented or collapsible containers
•
Large-volume, non-vented containers
•
Small-volume vials
•
60mL Luer syringes
Tip!
Baxter recommends using only sterile BD® 60 mL Luer syringes. If you plan to use a
different type, contact Baxter Technical Services. Refer to Getting Help on Page 22.
1.
Gather all the new ingredients.
Tip!
Baxter recommends using the Authorization Report to quickly identify the
ingredients needed for a specific configuration. For more information, refer to
Authorization Report on Page 167.
2.
Check that each source container has a barcode label attached.
Tip!
Baxter strongly recommends using the manufacturer’s barcode whenever possible.
NOTE:
For containers that are filled or diluted in the pharmacy, the Product Barcodes
Report can make labels available for printing. For more information, refer to Product
Barcodes Report on Page 178.