Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 44
53007410 Rev. C
Removing the Expired Tube Set and Expired Ingredients
CAUTION
Do not remove the valve set until you have removed all the source containers.
This precaution helps to prevent a dropped source container from damaging the
valve actuators.
WARNING
The only time the rotor should be rotated manually is under “Tube Set Will Be
Changed” or when the compounder is powered OFF. Turning the rotor at any other
time while the compounder is powered on will result in a fault state error message.
If an expired tube set and expired ingredients are already installed:
1.
Attach a calibration bag. Refer to Attaching the Calibration Bag on Page 73.
2.
Remove one source container from the vial rack or hanger, and turn the container right
side up.
3.
Lower the container below the height of the valve set, and allow the fluid in the inlet to
flow back into the source container.
4.
Repeat the previous steps for each source container.
5.
Press the tabs on the ends of the valve set, then lift to remove it.
Removing the valve set
NOTE:
The appearance of the valve set may differ from the example shown above.
6.
Turn the pump rotor counterclockwise to drain fluid from the common fluid pathway
into the calibration bag.
7.
Remove the calibration bag.
a.
Remove the bag’s fill port from the load cell’s fill port holder.
b.
Remove the bag from the load cell.
8.
Discard the valve set and the attached calibration bag.
9.
Discard the inlets, spikes and Ingredient(s) per the facility protocol.
Tip!
Baxter recommends cleaning the compounder before installing a new valve set.
Refer to CLEANING THE COMPOUNDER on Page 101, and follow your facility’s protocol.